dc.description.abstractNative heart valves that have limited functionality due to cardiovascular disease or congenital birth defects are commonly replaced by prosthetic heart valves. Bileaflet mechanical heart valves (BMHV) are the most commonly implanted valve design due to their long-term durability. However, their unnatural hemodynamics promote thrombosis and thromboembolic events. Clinical reports and in vitro experiments suggest that the thrombogenic complications in bileaflet valves are related to the stress imposed on blood by the valves during the closing phase. Additionally, animal and clinical studies have shown that BMHV in the aortic position demonstrate reduced failure rates compared to identical valves in the mitral position. The present study aimed to investigate the leakage, hinge, and near hinge flow fields of two BMHV under simulated physiologic aortic flow conditions and to compare these results with previous findings in the mitral position to better understand how the implant location influences the valve performance and the subsequent risk of blood damage. Two and three-component Laser Doppler Velocimetry techniques were used to quantify the velocity and turbulent shear stress fields in both the hinge and the upstream leakage flow regions. The study focused on the 23 mm St. Jude Medical Regent (SJM) and the 23 mm CarboMedics (CM) valves. Although they were tested under similar physiologic conditions, shape and location of the leakage jets were dependent on valve design. Nevertheless, turbulent shear stress levels recorded within all jets were well above the threshold shear stress for the onset of blood cell damage. Within the hinge region, the flow fields were complex and unsteady. The angulated hinge recess of the CM valve appeared to promote blood damage while the streamlined geometry of the SJM valve contributed to better washout of the hinge region. Animations of the velocity flow fields are given in QuickTime or MPEG format. Comparison of the present findings with previously published results for the mitral position suggests that the superior clinical results of the mechanical valves in the aortic position may be due to less severe leakage flow upon valve closure as well as to enhanced hinge washout during the forward flow phase.en_US
dc.format.extent522331 bytes
dc.format.extent617659 bytes
dc.format.extent615868 bytes
dc.format.extent602588 bytes
dc.format.extent457750 bytes
dc.format.extent578272 bytes
dc.format.extent535487 bytes
dc.format.extent534205 bytes
dc.format.extent419646 bytes
dc.format.extent611619 bytes
dc.format.extent925834 bytes
dc.format.extent1009631 bytes
dc.format.extent976930 bytes
dc.format.extent780180 bytes
dc.format.extent9091193 bytes
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/mpeg
dc.format.mimetypevideo/quicktime
dc.format.mimetypevideo/mpeg
dc.format.mimetypevideo/quicktime
dc.format.mimetypeapplication/pdf
dc.language.isoen_US
dc.publisherGeorgia Institute of Technologyen_US
dc.subjectLaser Doppler velocimetryen_US
dc.subjectThrombosis
dc.subjectHinge
dc.subjectImplant location
dc.subjectCarboMedics valve
dc.subjectSt. Jude Medical Regent valve
dc.subjectLeakage jets
dc.titleInfluence of the Implant Location on the Hinge and Leakage Flow Fields Through Bileaflet Mechanical Heart Valvesen_US
dc.typeThesisen_US
dc.description.degreeM.S.en_US
dc.contributor.departmentChemical Engineeringen_US
dc.description.advisorCommittee Chair: Yoganathan Ajit; Committee Member: Sambanis Athanassios; Committee Member: Sotiropoulos Fotisen_US